Field Story
The Gymnasium Where Nobody Heard the Fire Drill
A school gymnasium with bare concrete walls and a metal roof measured RT60 of 4.8 seconds at 500 Hz. Morning announcements were completely unintelligible, and the fire alarm notification system failed its annual STI test with a score of 0.31. After installing acoustic baffles on 40% of the ceiling, RT60 dropped to 1.6 seconds and STI improved to 0.58.
School GymnasiumImpulse Response Measurement per IEC 61672-1
TL;DR
IEC 61672-1 defines the electroacoustic performance of sound level meters used in measurement chains for impulse response capture. When using the interrupted noise method, the meter must comply with IEC 61672-1 for the decay recording. When using the integrated impulse response method, the standard defines the performance of the analog front-end (microphone through preamp through ADC) that captures the raw signal. SonaVyx implements IEC 61672-1 compliant digital processing and leverages the full 24-bit resolution of modern audio interfaces to maximize the dynamic range available for impulse response measurement.
IEC 61672-1 for IR Measurement
While IEC 61672-1 is a sound level meter standard, its specifications for frequency response, dynamic range, and linearity directly apply to the measurement hardware used for impulse response capture.
Instrument Classification
- Class 1 instruments are recommended for IR measurement in performance spaces per ISO 3382-1
- Class 2 instruments are acceptable for IR measurement in ordinary rooms per ISO 3382-2
- The classification affects allowable measurement uncertainty in the final results
- Digital implementations can achieve performance between Class 1 and Class 2 depending on the analog front-end
Frequency Response Requirements
- The measurement system must have flat response (within class tolerance) across the frequency range of interest
- For room acoustic measurements, this is typically 63 Hz to 8 kHz
- IEC 61672-1 Z-weighting tolerances apply to the capture chain
- Microphone calibration files can correct for known deviations
Dynamic Range and Linearity
IEC 61672-1 specifies the linear operating range of the meter. For impulse response measurement, you need the maximum linear range to capture both the peak of the excitation and the decay to the noise floor. The linearity range determines whether your IR is distortion-free.
Practical Implications
When measuring impulse responses with a browser-based tool like SonaVyx, the IEC 61672-1 compliance depends on the combination of the external microphone, audio interface, and the digital processing. The digital processing meets or exceeds the standard; the limiting factor is typically the analog components.
Common Mistakes
- Not matching the instrument class to the measurement standard requirements
- Overdriving the input, causing clipping that corrupts the impulse response
- Setting gain too low, reducing the usable dynamic range
- Not calibrating the microphone frequency response
SonaVyx Implementation
The SonaVyx IR tool uses IEC 61672-1 compliant digital processing at 48 kHz sample rate. Import microphone calibration files for the SPL meter to correct frequency response. Extract RT60 with the RT60 tool. Verify system linearity with the transfer function. Check the RTA for spectral overview. Visit learning modules for detailed setup procedures.
Standard Reference
IEC 61672-1:
Try It Now
Open this measurement tool in your browser — free, no download required.
Last updated: March 19, 2026