Field Story
Subwoofer Phase War at the Amphitheater
An outdoor amphitheater deployed four subwoofers in a line array, but seats in rows 12-18 heard almost no bass below 120 Hz. Transfer function measurement revealed a 22 dB null at the crossover frequency caused by phase cancellation between the main tops and subs. Correcting the sub delay by 4.2 ms and flipping polarity on the outer pair restored flat response across all seats.
Outdoor AmphitheaterPhase Analysis per AES-2id
TL;DR
Phase analysis is the heart of AES-2id system alignment. The phase response of the transfer function reveals timing relationships between frequency components that determine whether speakers sum constructively or destructively at the crossover. AES-2id practitioners use phase to align subwoofers with mains, match multiple speaker zones, and optimize crossover filter settings. SonaVyx transfer function displays unwrapped phase with group delay, providing the information needed to achieve the tight phase matching (within 45 degrees through the crossover region) that separates professional alignment from amateur EQ adjustments.
Phase in System Alignment
Phase analysis reveals the timing behavior of the system across frequency. While magnitude shows how loud each frequency is, phase shows when each frequency arrives. Both must be optimized for a well-tuned system.
Crossover Alignment
- At the crossover frequency, both drivers must be in phase for constructive summation
- Phase difference greater than 90 degrees causes destructive interference (dip at crossover)
- Target: less than 45 degrees phase difference through the crossover region
- Use delay and polarity adjustments to achieve phase alignment
Sub-Main Alignment
- Measure phase of subwoofer system alone at the crossover frequency
- Measure phase of main system alone at the same frequency
- Adjust sub delay until both phases match within 30 degrees at crossover
- Verify with combined measurement showing constructive summation
Group Delay
Group delay is the negative derivative of phase with respect to frequency. Constant group delay means all frequencies arrive at the same time. Large group delay variations indicate timing problems that cause smeared transient response.
Common Mistakes
- Ignoring phase and focusing only on magnitude response
- Using delay to fix a polarity problem (or vice versa)
- Not measuring phase at the crossover frequency specifically
- Confusing wrapped and unwrapped phase displays
SonaVyx Tools
Use phase analysis in the SonaVyx transfer function. Check polarity with the problem detector. Monitor levels with the SPL meter. View impulse timing with the IR tool. Get AI alignment recommendations from the diagnostic engine. Follow the PA tuning workflow.
Standard Reference
AES-2id:
Try It Now
Open this measurement tool in your browser — free, no download required.
Last updated: March 19, 2026