IEC 60268-16: STI Reporting and Compliance Documentation

|

TL;DR

IEC 60268-16 Clause 7 specifies reporting requirements for STI measurements. Reports must include: measurement method (direct STI, STIPA, or IR-derived), test signal type and level, ambient noise conditions, PA system settings (gain, EQ, zones), measurement positions (plan/drawing), equipment used with calibration status, per-position STI values, spatial average and standard deviation, and the applicable quality threshold. For EN 54-16 compliance, a zone-by-zone pass/fail matrix is required. The report should state whether the masking model was applied and which edition of the standard was used.

Minimum Reporting Requirements (Clause 7)

A compliant STI measurement report must document:

  • Method: Direct STI, STIPA, or STI from impulse response
  • Test signal: Type, level at the PA system input, and playback configuration
  • PA system state: Gain settings, EQ, zone configuration, AGC/limiter status
  • Ambient noise: LAeq during measurement, description of noise sources, whether HVAC was running
  • Equipment: Microphone model, analyzer model/software, calibration date
  • Positions: Plan drawing showing source and receiver locations with coordinates
  • Results: STI per position, spatial mean, standard deviation
  • Standard edition: IEC 60268-16 Ed.5 (2020) — results from different editions may not be comparable

EN 54-16 Compliance Format

For emergency voice alarm systems, the commissioning report typically includes:

ZonePositions measuredSTI minSTI maxSTI avgPass/Fail
Zone A - Lobby80.520.670.59PASS
Zone B - Corridor120.480.610.54PASS
Zone C - Car park60.410.530.47FAIL

Each zone is assessed independently. The minimum STI at any measurement position in the zone must meet the threshold (0.50 for EN 54-16), not just the average.

Common Reporting Mistakes

  • Reporting only the average: The worst-case position often matters more than the mean. Always report minimum, maximum, and mean.
  • Unrepresentative noise conditions: Measuring during quiet periods overstates real-world performance. Document the noise conditions clearly.
  • Wrong test signal level: If the PA gain was adjusted for measurement, the STI does not represent actual use conditions.
  • Missing masking correction: IR-derived STI without masking correction may overestimate intelligibility in noisy environments.
  • Mixing standard editions: Ed.4 and Ed.5 can differ by up to 0.05 STI due to the updated masking model.

Post-Measurement Actions

When positions fail to meet the target STI threshold:

  • Add speakers to improve coverage and direct-to-reverberant ratio
  • Add acoustic treatment to reduce reverberation time
  • Adjust speaker EQ to improve speech band energy (500 Hz – 4 kHz)
  • Add delay speakers to improve MTF at distant positions

SonaVyx Report Generation

The report generator creates STI measurement reports including per-position results, zone summaries, floor plan position mapping, and pass/fail assessment against your selected standard. Reports can be exported for client delivery. See measurement procedures for the workflow leading to the report.

Try It Now

Open this measurement tool in your browser — free, no download required.

Open Tool

Last updated: March 19, 2026