IEC 60268-16 STIPA: The Simplified STI Measurement Method

|

TL;DR

STIPA (Speech Transmission Index for Public Address systems) is a simplified single-number measurement defined in IEC 60268-16 Annex F. Instead of measuring 98 MTF values, STIPA uses a specially designed test signal with 2 modulation frequencies per octave band (14 total, per Table F.1), simultaneously modulating all 7 bands. A 15-second measurement captures the modulation depth at each frequency. STIPA correlates within ±0.03 of full STI for typical PA system conditions. It is the standard method for fire alarm voice evacuation system commissioning per EN 54-16.

Why STIPA?

Full STI measurement requires either a complete 98-point MTF grid or a high-quality impulse response. STIPA provides a practical alternative that can be measured in 15 seconds using a standardized test signal played through the PA system. This makes it ideal for large-scale commissioning where hundreds of measurement positions are needed.

Test Signal Design (Annex F, Table F.1)

The STIPA test signal consists of 7 simultaneous octave-band noise signals (125 Hz to 8 kHz), each amplitude-modulated at 2 specific frequencies. The modulation frequencies are chosen so that no two bands share the same pair, and all 14 of the standard modulation frequencies (0.63 to 12.5 Hz) are represented:

Band (Hz)Mod freq 1 (Hz)Mod freq 2 (Hz)
1251.005.00
2500.636.30
5000.808.00
10001.2510.0
20001.6012.5
40002.003.15
80002.504.00

The modulation depth is 0.55 (-5.2 dB relative to full modulation). Each band is independently modulated so crosstalk between bands can be detected.

Measurement Procedure

  1. Play the STIPA test signal through the PA system at normal operating level
  2. Record 15 seconds of the received signal at the measurement position
  3. Bandpass filter into 7 octave bands
  4. Extract the modulation depth at each of the 14 frequencies
  5. Compute MTF as the ratio of received to transmitted modulation depth
  6. Follow the standard STI calculation pipeline

Accuracy and Limitations

STIPA agrees with full STI within ±0.03 for most PA system conditions. However, accuracy degrades when:

  • Strong discrete echoes (flutter echo) create MTF peaks at specific modulation frequencies not sampled by STIPA
  • Nonlinear distortion (clipping) affects specific bands asymmetrically
  • Very long reverberation times (T > 4s) where the 15-second measurement may be insufficient

For these edge cases, measurement from impulse response or extended-duration STIPA (3× averaged) provides better accuracy.

EN 54-16 Application

The European fire alarm voice evacuation standard EN 54-16 mandates STIPA measurement at every listener position during system commissioning. The minimum requirement is STI ≥ 0.50 (Fair). Many specifiers require ≥ 0.55 or ≥ 0.60 to provide a safety margin. See measurement procedures for equipment requirements.

SonaVyx STIPA

The STI tool generates the standard STIPA test signal and measures the received modulation transfer function. Quick mode uses a single 15-second capture; Pro mode averages three captures for improved reliability. Results include the qualitative rating and can be included in compliance reports.

Try It Now

Open this measurement tool in your browser — free, no download required.

Open Tool

Last updated: March 19, 2026